Analytical chemistry, Jan 4, 2018
Lateral flow assays (LFAs) are widely used for yes/no detection of analytes, but they are not wel... more Lateral flow assays (LFAs) are widely used for yes/no detection of analytes, but they are not well-suited for quantification. We show that the sensitivity of the test line in a lateral flow assay can be tuned to appear at a specific sample concentration by varying the density of capture molecules at the test line and that when test lines tuned for different responses are combined into a single test strip, lines appear at specific thresholds of sample concentration. We also developed a model based on mass-action kinetics that accurately described test line signal and shape over a wide matrix of capture molecules and sample concentrations in single-line strips. The model was used to design a three-line test strip with lines designed to appear at logarithmically spaced sample concentrations, and the experiments showed a remarkable match to predictions. The response of this "graded ladder bar" format is due to the effect of test line concentration on capture efficiency at each...