Advertisement
Home > Columns > Column: Column Watch
News | Articles | April 6, 2026 (Updated: April 7, 2026)
Slalom Chromatography (Part 1): From Gels to Modern UHPLC
Slalom Chromatography
Author(s) Fabrice Gritti , Jamuna Vaishnav , Balasubrahmanyam Addepalli
+2 More
https://doi.org/10.56530/lcgc.int.dk3184p9
Listen
0:00 / 45:02
Key Takeaways
Early SC separations inverted SEC/HDC elution order under non-adsorptive conditions,
motivating a retention concept based on DNA reorientation and hydrodynamic
interactions within packed beds.
Abandonment reflected missing pH-stable, mechanically robust, low-adsorption
packings, bio-compatible hardware, complex mobile-phase requirements, and absent
physicochemical validation of retention and mass-transfer models.
SHOW MORE
Part 1 of this two-part article series reviews the history of slalom chromatography (SC), the rationale
for its recent revival, and the updated mechanistic framework of this technique. Rooted in the
fundamental physical properties of deoxyribonucleic acid/ribonucleic acid (DNA/RNA) biopolymers and
laminar flow dynamics in packed chromatographic beds, SC is highlighted for its most promising
applications in cell and gene therapy. Recent two-year experimental investigations have revealed that
SC operates through an out‑of‑equilibrium mechanism, coupling the entropic elasticity of
double‑stranded deoxyribonucleic acid/ribonucleic acid (dsDNA/RNA) with the extension and shear
forces generated within the interparticle spaces of ultrahigh-pressure liquid chromatography (UHPLC)
columns. In contrast to gel permeation chromatography (GPC) or hydrodynamic chromatography
(HDC), SC functions at high speed and uniquely elutes smaller dsDNA fragments first, followed by
progressively larger ones. Unlike agarose gel electrophoresis (AGE), SC achieves separations of large
DNAs in less than three minutes, with nearly twice the resolution power of typical AGE. These findings
paved the way for the recent design of a new SC column intended for use in cell and gene therapy as
an alternative to standard AGE. Further applications of this new SC column within cell and gene
therapy workflows will be discussed in Part 2.
In the 1980s, the most widely used technique for separating, quantifying, and preparing
double-stranded deoxyribonucleic acid (dsDNA) fragments was agarose gel electrophoresis
(AGE).1 However, the separation of large DNAs above 20 kbp was not achievable with
conventional AGE until pulsed‑field gel electrophoresis (PFGE) was introduced.2,3 The main
limitations of AGE and PFGE are their procedural complexity (gel preparation, voltage
supply, staining, and DNA detection) and their long analysis times, approximately 2 h for
AGE and more than 10 h for PFGE.
To address these issues, in 1988, Boyes4 and Kasai5 tested the performance of gel
permeation chromatography (GPC)—a technique widely used for protein separation—for
large DNA separations. Remarkably, both research groups independently demonstrated
that large dsDNA fragments could indeed be successfully separated using conventional gel
permeation chromatography supports. Kasai termed this new separation mode slalom
chromatography (SC), and the name was officially born. While this breakthrough opened the
door to the development of new columns for the rapid characterization of large DNA
samples, the technology failed to gain significant attention. Between 1988 and 2004, no
more than 15 original research articles—fewer than one per year—were published on both
the experimental6–16 and theoretical17–19 fronts. After this period, SC was abandoned and
no longer considered as a potential alternative to AGE. The reasons why SC was not fully
recognized or supported as a powerful separation technique over the past two decades
have not been clearly addressed in the literature. This merits attention and better
understanding given all the significant improvements in column technologies following the
introduction of ultrahigh-pressure liquid chromatography (UHPLC).
In Part 1 of this series, we first revisit the historical context of SC, explaining why it was not
adopted as a new promising separation technique until 2023, when Waters Corporation
invested research resources into the commercialization of the first SC column. This part
specifically highlights the extensive experimental investigations undertaken to unravel the
hidden retention and mass‑transfer mechanism in SC. It also demonstrates how the unique
retention and mass‑transfer properties of SC are deeply rooted in the physics of single DNA
biopolymers and fluid hydrodynamics in a packed chromatographic column, and why SC
can now compete well with the current gold standard in DNA separation, AGE, based on
the quality-by-design (QbD) of a new SC column released in 2025. The main applications in
cell and gene therapy workflow of this new SC column will be discussed in Part 2.
Materials and Methods
The Optima-grade solvents used in this work were water and acetonitrile from Fisher
Scientific. Phosphate buffered saline (PBS) tablets, sodium monophosphate salt (NaH2PO3),
disodium phosphate salt (Na2HPO3), 10X Tris Acetate EDTA (TAE), sucrose, and cytosine
were also acquired from Fisher Scientific. The three dsDNA sample mixtures, Lambda DNAHind III Digest (2.0, 2.3-, 4.4-, 6.6-, 9.4-, and 23.1 kbp dsDNAs), Lambda DNA-BstEII Digest
(1.264-, 1.371-, 1.929-, 2.323-, 3.675-, 4.324-, 4.822-, 5.686-, 6.369-, 7.242-, and 8.454-kbp
dsDNAs), and Lambda DNA-Mono Cut Mix (1.5-, 10.0-, 15.0-, 17.0-, 24.0-, 24.5-, 29.9-, 33.5-,
38.4-, and 48.5-kbp dsDNAs), were procured from New England BioLabs Inc.
Three homemade research prototype columns with MaxPeak Premier Technology column
hardware (4.6 mm internal diameter (i.d). x 300 mm long) were used in this work. These
research prototype columns were packed with monodisperse non-porous 2.6-µm silica
particles obtained on-site (Waters), porous 1.7-, 2.4-, 5.4-, and 11.3-µm BEH125 particles
produced on-site (Waters), and with polydisperse non-porous 3.0-µm silica particles
purchased from Glantreo. The restrictor capillaries (40 µm i.d. x 30 inch, 40 µm i.d. x 24 inch,
and 40 µm i.d. x 16 inch) were homemade and connected to the column outlet to increase
the average pressure along the column.
The new column dedicated to SC applications and used in this work is the 4.6 x 300 mm,
2.5-µm Waters GTxResolve 250 Å Slalom Column, MaxPeak Premier Technology. It is packed
with 2.5 µm particles in a metal column hardware modified with a hydrophilic vapor-
deposited surface. The particle surface is made inert to DNA after modification with a
polyethylene oxide-bonding treatment.
All the chromatograms were recorded on an Acquity UPLC I-Class Plus System (Waters). It is
equipped with a binary pump with solvent selection valves, a flow-through needle (FTN)
autosampler, an external one-column heater module (embedding 4.6 x 300 mm columns),
active solvent pre-heaters, left and right column selection valves (each with nine ports and
eight positions), and a tunable wavelength UV–vis detection system (500 nL cell). The
sample loop (125-µm i.d.), the needle seat capillary (75-µm i.d.), and the active pre-heater
tubes (75-µm i.d.) were all surface-modified with hydrophilic hybrid organic-inorganic silica.
The system is controlled by EmpowerSoftware 3.0 (Waters) and can be operated up to a
maximum pressure of 15,000 psi at 1 mL/min.
A Brief History of Slalom Chromatography Trials (1988–2004)
In 1988, Jun Hirabayashi and Ken‑ichi Kasai of the Faculty of Pharmaceutical Sciences in
Sagamiko, Japan, together with American researchers Barry Boyes, Douglas Walker, and
Patrick McGeer from the Department of Psychiatry at the University of British Columbia in
Vancouver, Canada, independently discovered that double‑stranded DNA (dsDNA)
fragments with molecular weights exceeding 2 MDa could be separated using conventional
size‑exclusion chromatography (SEC) columns. The Japanese group employed porous SEC
particles of 5 µm and 9 µm with average pore diameters of 250 Å,5 while the Canadian
group used 4-µm particles with the same mesopore size.4 These packing materials were
originally designed for the separation of large biomolecules such as monoclonal antibodies,
typically around 100 Å in size.
Remarkably, under non‑adsorptive conditions, both groups observed that the elution order
of dsDNA fragments, ranging from 2 kbp to 23.1 kbp in length and corresponding to
hydrodynamic diameters of approximately 0.1 µm to 0.5 µm, was the exact opposite of the
conventional exclusion mechanism normally associated with SEC or hydrodynamic
chromatography (HDC). The very first SC chromatograms ever recorded are shown in Figure
1. At the time, neither group could provide a physicochemical explanation for this
unexpected phenomenon.
Figure 1. (a) First separation of large dsDNA fragments (HindIII digest of a 48.5 kbp λDNA, from 2 kbp to 23.1 kbp) reported by Boyes et al.1 in 1988, performed on a GPC
DuPont Bioseries GF-250 column (9.4 mm i.d. x 250 mm) packed with 4 μm silica spheres
modified with a hydrophilic diol-bonded phase. Mobile phase: 50 mM Tris buffer, 0.5 mM
EDTA, pH = 7.5. Flow rate: 0.25 mL/min. Room temperature. (b) First separation of the
same large dsDNA fragments shown in the left panel, reported by Kasai et al 2 in 1988,
performed on a GPC Asahipak GS-510 column (7.6 mm i.d. x 500 mm) packed with 9 μm
porous polymer-based particles. Mobile phase: 100 mM Tris-HCl buffer, 0.2 M NaCl,
pH=7.5. Flow rate: 0.60 mL/min. Room temperature.
The origins of SC are firmly historical. The term was first introduced, arbitrarily, by
Hirabayashi and Kasai as a new mode of retention and separation in liquid chromatography
(LC) in their 1989 publication, Size‑Dependent Chromatographic Separation of
Double‑Stranded DNA Which Is Not Based on Gel Permeation Mode.6 The authors intuitively
envisioned a DNA polymer behaving like a skier, “slaloming” or repeatedly turning left and
right around the packed particles during its migration through the column. Based on this
analogy, they proposed a separation mechanism governed by the differential reorientation
times of dsDNA molecules: longer fragments would require more time to reorient than
shorter ones, thereby explaining the observed order of elution.
Over the following 16 years, until the technique was effectively abandoned in 2004,
Hirabayashi and Kasai remained the only researchers to conduct significant experimental
investigations aimed at elucidating the separation mechanism of SC. Experimental studies
published in the 1990s7–9 and in 200010 demonstrated that DNA retardation was primarily
influenced by DNA length, particle size, flow rate, and solvent viscosity. However, they
ultimately concluded that the precise retention mechanism remained inexplicable. During
the early 2000s, similar experimental investigations were performed by Peyrin and
Guillaume11–16 before a handful of theoretical models could be proposed.17–19 Yet, the
mathematical frameworks proposed lacked robust physicochemical validation, as they were
not supported by direct observations of large dsDNA molecules migrating through packed
beds, a limitation highlighted in a recent review by Gritti.20
After 2004, coinciding with the rise of ultrahigh‑performance liquid chromatography
(UHPLC) and the introduction of sub-2-µm particles, experimental interest in SC declined
sharply. Aside from a few retrospective reviews,21–24 no new experimental studies or reports
appeared in the scientific literature. The technique was effectively abandoned due to
several critical limitations: the lack of suitable packing materials with high pH stability, low
non‑specific adsorption, and adequate mechanical strength; bio‑compatible column
hardware; the requirement for complex mobile phases containing salts, chelating agents
such as ethylenediaminetetraacetic (EDTA), and organic solvents; the absence of a clear
understanding of the SC retention mechanism; and a limited application market.
Additionally, several alternative separation techniques were already available, though
primarily suited for smaller DNA fragments. As shown in Figure 2, analytical techniques
such as ion pairing reversed‑phase (IPRP)–LC, anion exchange chromatography (AEX),
hydrophobic interaction chromatography (HIC), and size exclusion chromatography (SEC)
were commonly employed for sub‑MDa DNA. For larger DNA molecules, techniques
including hydrodynamic chromatography (HDC), capillary electrophoresis (CE), AGE, PFGE,
analytical ultracentrifugation (AUC), and asymmetric flow field‑flow fractionation (AF4) were
widely adopted despite their intrinsic limitations in terms of simplicity, speed, resolution,
and accuracy.
Figure 2. List of existing analytical separation techniques (bottom) commonly used for the
characterization of various biopharmaceutical modalities (top) as a function of molecular
size. These modalities range from small molecules to proteins, adeno associated viruses
(AAVs), exosomes, virus like particles (VLPs), lipid nanoparticles (LNPs), DNA, and cells. In
green, emphasis is placed on the separation of large dsDNA fragments ranging from 2 kbp
to 50 kbp, or with hydrodynamic diameters from 0.1 μm to 0.7 μm, for which classical
adsorption techniques such as RP IP, AEX, HIC, and HILIC are not suitable due to limited
resolution. FFF lacks sufficient resolution, while HDC and AUC are both complex
separation techniques. Currently, AGE and PFGE represent the gold standard methods for
separating large DNA up to 10 Mbp; however, they suffer from extremely long analysis
times. This limitation justifies the revival of a simple, fast, high-resolution, and accurate SC
separation technique.20
The Slalom Chromatography Revival
Renewed interest in this legacy separation technology has largely been driven by growing
demand within the biopharmaceutical industry in the late 2010s to improve the
characterization of large DNA and ribonucleic acid (RNA) biopolymers, which exceed 2 MDa
in molecular weight and are often highly heterogeneous in both composition and structure.
Accurate determination of size and sequence is essential for a wide range of applications,
including cell and gene therapy, for example, messenger ribonucleic acid (mRNA) vaccines,
DNA restriction mapping, directed evolution via protein selection, and clustered regularly
interspaced short palindromic repeats-Cas9 (CRISPR‑Cas9) gene editing. These advances
have underscored the need for robust analytical tools capable of resolving complex nucleic
acid mixtures with high precision.25
In the early 2020s, the market was pressuring manufacturers to develop new DNA/RNA
analytical techniques that were simultaneously simple, fast, high-resolution, accurate, and
reproducible.26 Under these conditions, SC emerged as a highly suitable candidate for LC
manufacturers compared with the separation technologies listed in Figure 2. A recent
review20 highlights the main limitations of IPRP, AEX, hydrophobic interaction
chromatography (HIC), hydrophilic interaction chromatography (HILIC), and size exclusion
chromatography (SEC) in terms of resolution for DNA sizes greater than a few kbp. Of the
others, HDC is complex and restricted to nanoflow, narrow open capillaries, and laser-
induced fluorescence detection; AUC requires complex data processing; and AF4 offers only
limited resolution. AGE and PFGE remain the most widely adopted techniques for achieving
high-resolution separation of DNA mixtures larger than a few kbp but their analytical
throughput is slow, requiring one to two hours to complete a single gel.
SC emerged as a suitable technique in the early 2020s because pH-stable, non-adsorptive,
and mechanically robust sub-3-µm particles were already being manufactured for SEC,
designed for the characterization of various peptides, proteins, and monoclonal antibodies.
At the same time, bio-inert hardware was being developed to mitigate strong metal–
analyte interactions that can cause poor peak shapes and irreversible analyte losses.27–30
These attributes are essential for performing high-resolution separations and handling the
many phosphate groups present in DNA/RNA modalities.
Nevertheless, in the absence of a clearly established separation mechanism in SC, several
critical and fundamental questions remained unanswered regarding the design of the novel
SC columns: which particle size, flow rate, temperature, operating column pressure, column
length, and buffer/salt concentrations are truly required to achieve simple, fast, highresolution, and reproducible separations of large DNA/RNA biopolymers by SC?
Slalom Chromatography Retention and Efficiency Properties:
The Road To Victory (2023–2024)
To address these fundamental questions, the immediate course of action was to perform
systematic experiments by considering seven different experimental parameters, varying
one at a time while keeping all others constant. Supplementary Figures S1a–S1g (to be
found after the references) summarize these early empirical investigations conducted prior
to the commercialization of the first GTxResolve Slalom Chromatography Column. The
intrinsic effects of each parameter were carefully analyzed: hydrostatic pressure P
(Supplementary Figure S1a, by connecting capillary restrictors at the column outlet, with
average column pressures ranging from 900 psi to 7080 psi); temperature T (Supplementary
Figure S1b, ranging from 25 oC to 90 oC); mobile phase viscosity ƞ (Supplementary Figure
S1c, by adding sucrose up to 40% by weight in the eluent, ranging from 0.9 cP to 5.2 cP);
average interstitial linear velocity u (Supplementary Figure S1d, by increasing flow rate from
0.15 mL/min to 1.10 mL/min through a 4.6-mm-i.d. column); ionic strength I of the mobile
phase (Supplementary Figure S1e, from 0.001×PBS to 1×PBS pH 7.4 buffer concentration,
with both silica particles and DNA negatively charged); particle size dp (Supplementary
Figure S1f, ranging from 1.8 µm to 11.3 µm, same strain force); and surface chemistry of the
column hardware (Supplementary Figure S1g, bare stainless steel vs. hybrid surface
technology [HST]–modified hardware). The empirical conclusions related to the dsDNA
retention mechanism in SC became clear and are summarized as follows20, 31–33:
1. DNA retention is independent of hydrostatic pressure under non-adsorbed conditions.
2. DNA retention decreases with increasing temperature as a result of reduced solvent
viscosity, which diminishes extensional and shear forces in between the packed
particles.
3. DNA retention increases with higher sucrose concentrations as solvent viscosity rises.
4. DNA retention volume increases with higher linear velocity as extensional and shear
forces intensify and average DNA extension length (Lext) increases.
5. DNA retention decreases with lower ionic strength, owing to stronger electrostatic
repulsions between the negatively charged silica surface (pH ~ 8) and the phosphate
backbone of DNA polymer chains.
6. DNA retention decreases at constant extensional/shear force with increasing particle
diameter. When the ratio of DNA extension length to particle diameter Lext/dp falls
below about 25%, retention is governed by a pure HDC exclusion mechanism. When
this ratio exceeds one, a pure SC retention mechanism applies, and the retention factor
increases linearly with the average extension length best estimated from the known
force–extension curve of a single DNA polymer chain.34–36,20,31
7. DNA peak shape is deteriorated and DNA mass is lost when biopolymers are exposed
to stainless-steel surfaces compared with hybrid surface technology-modified
hardware.27–30
It required nearly two years to plan, collect, and analyze these empirical data, and to
provide a consistent interpretation throughout. Figure 3 shows the variation of the zone
retention factor k1 (the reference hold-up time is the external interparticle void time) of
dsDNAs of sizes ranging from 1.5 kbp to 48.5 kbp measured on the GTxResolve 250Å
Slalom Column, MaxPeak Premier Technology (4.6 x 300 mm, 2.5 µm when run at 25 oC
using a 10XTAE buffer pH 8.0 mobile phase and a series of flow rates from 0.15-, 0.30-,
0.50-, 0.70-, 0.90- to 1.10 mL/min.37 When the average extension length exceeds the
particle diameter (dp), these results demonstrated convincingly that the retention of
dsDNAs (k1 > 0.2) of varying sizes is governed by a unique extension/shear stress
parameter, which is written:
Equation 1
From a practical viewpoint, the experimental data indicate that the retention of DNA is
unchanged as long as the stress parameter τ remains constant, irrespective of temperature,
flow rate, and particle diameter used.
In contrast, Lext when is shorter than approximately 25% of dp, the DNA biopolymer is too
small and is simply excluded according to a conventional HDC retention mechanism (k1 <
0). For intermediate extension lengths (0.25dp < Lext < dp), a mixed HDC–SC retention
mechanism occurs and the selectivity between critical pairs of linear dsDNAs reaches a
maximum for DNA sizes around 6–7 kbp.31,32 As the extended length of the DNA increases,
it becomes less excluded from the particle surface (hence the HDC retention factor is less
negative) and the probability that the DNA chain is in proximity of the particle surface
increases (hence the SC retardation increases).
Supplementary Figure S2 perfectly illustrates this mixed HDC (exclusion)–SC (retardation)
retention mechanism and presents the same plots as in Figure 3, obtained at a constant
flow rate of 1 mL/min, but under three different ionic strengths: 4-, 40-, and 400-mM TAE.
As expected, increasing the buffer concentration increases DNA retention because this
more effectively screens the negative charges on the silica particle surface, allowing the
extended DNA chains to be less excluded, approach closer to the particle surface, be more
stretched, and experience more local velocities approaching zero.
Figure 3. Retention properties of the 250 Å. BEH GTxResolve Slalom Column (4.6 mm i.d. x
300 mm, 2.5- μm) as a function of the size of linear dsDNA fragments (ranging from 1.5
kbp to 50 kbp) across seven applied flow rates: 0.15-, 0.3-, 0.5-, 0.7-, 0.9-, and 1.1mL/min.
Mobile phase: 10xTAE buffer, pH 8. Room temperature. Each data point corresponds to the
elution of dsDNA fragments present in three λ-DNA digest mixtures (HindIII, BstEII, and
Mono Cut Mix). Three distinct retention regimes are observed depending on the ratio of
the extended length of the dsDNA fragments to the particle diameter: the HDC separation
mechanism applies (k1 < 0) when the ratio falls below 25%; the SC retention mechanism
applies (k1) > 0) when the ratio exceeds 1. Most remarkably, selectivity is highest when a
mixed HDC–SC retention mechanism occurs at intermediate values of this size ratio. For
the GTxResolve Slalom Column, maximum selectivity is expected for DNA fragments
around 7 kbp.37
Interestingly, as shown in Figure 4, and from a purely empirical perspective, the efficiency of
the recorded DNA peaks appears to be a uniquely decreasing function of the retention
factor (k1), regardless of the choice among the many possible experimental parameters. In
practice, it is generally recommended to maintain k1 below 0.3 to maximize the resolution
power of the slalom column, particularly around k1 ~ 0.32
Figure 4. Efficiency properties of the recently commercialized 250 Å BEH GTxResolve
Slalom Column (4.6 mm i.d. x 300 mm, 2.5 μm) as a function of the retention factor (k1) of
linear dsDNA fragments under varying experimental conditions of temperature (5–70 oC),
flow rates (0.1–1.4 mL/min), and TAE buffer concentration (4–400 mM). Mobile phase: TAE
buffer, pH 8. Each data point corresponds to the elution of dsDNA fragments present in
three l-DNA digest mixtures (HindIII, BstEII, and Mono Cut Mix). A strong empirical
correlation was observed between column efficiency and DNA retention, explained by the
change in relaxation time (tR) required for dsDNA fragments to transition from their
extended, retained, non equilibrium conformation to their random coil, free, equilibrium
conformation. Maximum resolution factors, reflecting the optimal trade-off between
selectivity (Figure 3) and plate counts (this figure), are consistently observed at k1 ~ 0.72.32
The translation of this empirically determined separation mechanism into a real-life
application is illustrated in Figure 5 (DNA restriction mapping), which shows the highresolution separation of 13 linear dsDNA fragments (sizes from 1.3 kbp to 14.4 kbp
including the cohesive ends assembly of the 5.686 and 8.454 fragments as revealed in
Supplementary Figure S1d) generated by the BstII digest of λ-DNA (48.5 kbp linear dsDNA).
Figure 5. Illustration of the separation performance of a 250 Å BEH GTxResolve Slalom
Column (4.6 mm i.d. x 300 mm, 2.5 μm) for DNA restriction mapping following injection
of 2 mL of the BstEII λ-DNA digest containing 12 dsDNA fragments, as indicated in the
chromatogram (fragment sizes shown in blue, kbp units). Notably, the 5.686 and 8.454
kbp fragments carry complementary 5’ strand overhang (12 bases) that hybridize to form
a new 13th fragment of 14.14 kbp (fragment sizes shown in green), eluting at 2.55 min.
Mobile phase: 10xTAE buffer, pH 8. Flow rate: 1.1 mL/min. Room temperature. In
agreement with Figures 3 and 4, the peak capacity of the SC column is maximal for the
separation of 4–9 kbp DNA fragments, reaching 11 in less than 2.5 min (combined HDC-
SC separation mechanism, -0.15 < k1 < 0.20). For comparison, the peak capacity
measured by AGE is only 7, with gel runs lasting up to 1 h (see top right corner insert).
Separation is extremely limited for DNA fragments smaller than 2.5 kbp (full HDC-based
separation mechanism, k1 < -0.20) and moderate for fragments larger than 11 kbp (full
SC separation mechanism, k1 > 0.25).
These results confirm that column efficiency decreases with increasing retention factor, as
evidenced by broader peak widths, and that resolution is maximized for DNA fragments
around 6–7 kbp compared with lower resolution factors observed for fragments between
1.3–3.7 kbp or above 10 kbp. Overall, an excellent peak capacity of 11 was achieved for the
elution of 4.3 kbp to 8.5 kbp linear dsDNA fragments in less than 2.5 min. By contrast, the
insert in Figure 6 shows the AGE separation (1% agarose gel), which required 1.5 h to
complete and yielded a peak capacity of only 7.
Figure 6. Changes in the conformation of a linear dsDNA fragment as a function of the
force applied to one end while the other end is fixed. The force range extends from less
than 10 fN (dsDNA random coil conformation) to 1 nN (ssDNA breakage). The linear
dsDNA is first uncoiled through weak entropic elasticity (persistence length ~ 500 Å),
reaching up to 80% of its contour length at a force of 2 pN. It is then slightly extended via
strong enthalpic elasticity up to ~65 pN, at which point the double helix structure is lost
and the molecule becomes overstretched. The resulting ssDNA strands are further
stretched enthalpically (strong forces applied) until the phosphodiester bonds break down
at ~1 nN (Bouchiat). Forces generated in the SC column never exceed 2 pN.
Immersion in the Physics of Slalom Chromatography
Figure 6 shows the changes in the configuration of a single linear dsDNA chain upon the
application of a force at one end while the other end is fixed. When the applied force is
smaller than 10 fN (1 fN=10-15Newton), the linear DNA still adopts its equilibrium random
coil conformation. For example, the hydrodynamic diameter of a 48.5 kbp λ-DNA is as large
as 0.7 µm , owing to the long persistence length (a measure of chain rigidity) of dsDNA,
approximately 500 Å. dsDNA is a semi-flexible biopolymer compared to single-stranded (ss)
DNA, which has a much lower persistence length of only about 10–20 Å.
When the applied force ranges from 10 fN to about 2 pN (1 pN=10-12 Newton), the dsDNA
coil is progressively undone until its extended length reaches the contour length. This
regime is governed by the entropic elasticity (low spring constant) of dsDNA, which is
inversely proportional to its persistence length.34–36 For example, single-stranded
deoxyribonucleic acid (ssDNA) and mRNA coils cannot be undone within this force range
because their persistence lengths are too short. From 2 pN to 60 pN, the extended length
of dsDNA increases only slightly, owing to the enthalpic elasticity (high spring constant) of
the unfolded DNA chain. Remarkably, the dsDNA length extends by approximately 90% as
it overstretches and loses its double-helix structure at a constant force of about 65 pN. For
forces greater than 65 pN, the ssDNA is extended according to its own enthalpic elasticity
until the weakest phosphodiester bond typically breaks at a force close to 1 nN (1 pN=10-9
Newton).
The question asked was then to evaluate the force involved in the interparticle volume of
SC columns. Considering a 4.6-mm-i.d. column packed with 2.5-µm particles and operated
at 1 mL/min, the force applied to 2.5 kbp and 50 kbp ranges typically from 100 fN to 2 pN,
that is, forces are sufficient to extend significantly such dsDNAs from 50% up to 80% of
their maximum contour lengths.20,31 The right panel in Figure 7 shows the relationship
between the force applied and the extension of 48.5 kbp linear λ-DNA as reported by
Bouchiat.36 The retention and mass transfer mechanism is then fully interpreted by the coilto-stretch transition of the linear dsDNA.
Figure 7. Physics-based explanation and description of the separation mechanism in SC:
(Top left panel) Evidence of local velocity gradients in the interparticle spaces of random
sphere packings obtained from numerical simulations of the Navier–Stokes equations in
fluid dynamics. These gradients generate a combination of extension and shear rates that
stretch the dsDNA biopolymer. (Bottom left panel) Direct visualization of the coiled and
extended conformations, as well as the coil–stretch transition, of a large linear 166 kbp
dsDNA polymer migrating through spaces defined by 10 mm i.d. pillars arranged in a
square array (Strom). (Right panel) Experimental data (open circles) showing the
relationship between applied force and extension length of a linear 48.5 kbp λ-DNA
biopolymer (Bouchiat). LC forces are sufficient to significantly stretch the DNA biopolymer.
The left panel in Figure 7 shows the calculated velocity profile of the mobile phase across a
cross‑section of a packed column and the two configurations of a large 166 kbp dsDNA
polymer migrating between 10-μm-diameter pillars arranged in a cubic array.38 The left
panel reveals that under laminar flow conditions, velocity gradients occur at many locations,
generating continuous extension forces (along the flow direction) and shear forces
(perpendicular to the flow direction) acting on the analytes. When the DNA polymer is
located far from the particle surface or near the center of a flow‑through channel, the
extension and shear rates are minimal, and the dsDNA relaxes and retains its coiled
configuration. It relaxes into an equilibrium “free” state and migrates rapidly at the local
velocity of the mobile phase. In contrast, when a fraction of the DNA polymer chain is
positioned close to the particle surface or within constricted regions of the separation
medium, the extension and shear forces are maximum and stretch the DNA chain. In this
non-equilibrium “retained” state, the DNA migrates at a slower velocity even though no
physisorption occurs. This situation is analogous to an adsorption–desorption process,
which effectively describes the DNA retention and efficiency observed. Retention is
governed by the size or length of dsDNA through its probability of being exposed to the
highest extension and shear forces, while efficiency is determined by the effective relaxation
times between the “retained” and “free” states.39 This resolves the mystery noted by Boyes
and provides a complete fundamental microscopic interpretation of the retention and
mass‑transfer mechanism in SC, an explanation that Kasai and others had been seeking for
more than 25 years.
Conclusion
In part 1, the history, resurrection, and retention and mass transfer mechanism of SC has
explained why this separation technique was rapidly abandoned in the early 2000s,
remained dormant for two decades, and was revived in the early 2020s. The combination of
strong demand from the biopharmaceutical industry to improve the characterization of
large DNA/RNA molecules, the absence of a clearly established retention and mass‑transfer
mechanism, and the availability of the latest high‑performance UHPLC particles (high‑class
purity type A, sub‑3-µm in size, free of non‑specific adsorption sites, with HST‑modified
stainless‑steel hardware) and systems (HST‑modified) propelled intensive empirical and
fundamental investigations in DNA separation, ultimately leading to the design of the first
commercialized slalom column in the history of chromatography.
What has been learned over the last three years now provides a clear explanation for the
early, previously unexplained, observations made by Boyes and Kasai regarding the
separation and efficiency of SC columns. This physics mechanism is based on the slow coilto-stretch transition of dsDNA biopolymers within the interparticle volume of a randomly
packed column. Retention occurs only for dsDNAs/dsRNAs, owing to their long persistence
length and weak entropic elasticity, whereas ssDNA and mRNA are excluded due to their
too short persistence lengths. Additionally, the contour length of retained dsDNA/dsRNA
must exceed approximately 25% of the particle diameter. Retention should not be too high,
as overly large dsDNA/dsRNAs exhibit slow relaxation times, causing extremely wide peaks,
let alone DNA trapping in the column. Consequently, the retention of 3–25 kbp dsDNA was
found to be optimal on a 4.6 × 300 mm column (bio‑inert HST‑modified hardware) packed
with 2.5-µm polyethylene oxide‑surface‑modified hybrid BEH particles. The internal particle
porosity (250 Å) was selected to enhance separation of the targeted dsDNA/dsRNA
analytes from smaller buffer molecules, proteins, enzymes, and oligomers, which are more
strongly retained according to a SEC retention mechanism.
In Part 2, we will illustrate real‑world applications of the new SC column in the field of cell
and gene therapies and compare its performance to that of the conventional AGE
separation technique. This includes characterization of plasmids (supercoiled vs. linear) and
quantification of dsRNAs in mRNA vaccines prepared by in vitro transcription (IVT). Focus
will be given on the practical advantages of the SC column, that is, simplicity, speed,
resolution, and analytical precision, for determining the length and mass of injected
dsDNA/dsRNA. SC’s overall performance will be benchmarked against the current gold
standard, AGE, to provide the biochemists and biologists with new options to enhance and
expedite their daily laboratory work.
Acknowledgments
Waters, MaxPeak, BEH, GTxResolve, Acquity, UPLC, and Empower are trademarks of Waters
Corporation or its affiliates. Optima is a trademark of Fisher Scientific Company LLC.
References
1. Maniatis, T. A.; Fritsch, E. F.; Sambrook, J. Molecular Cloning: A Laboratory Manual; Cold
Spring Harbor Laboratory Press, 1982.
2. Schwartz, D. C.; Cantor, C. R. Separation of Yeast Chromosome‑Sized DNAs by Pulsed
Field Gradient Gel Electrophoresis. Cell 1984, 37, 67–75. DOI: 10.1016/00928674(84)90301-5
3. Carle, G. F.; Olson, M. V. Separation of Chromosomal DNA Molecules from Yeast by
Orthogonal‑Field‑Alternation Gel Electrophoresis. Nucleic Acids Res 1984, 12, 5647–
5664. DOI: 10.1093/nar/12.14.5647
4. Boyes, B. E.; Walker, D. G.; McGeer, P. L. Separation of Large DNA Restriction Fragments
on a Size‑Exclusion Column by a Nonideal Mechanism. Anal Biochem 1988, 170(1),
127–134. DOI: 10.1016/0003-2697(88)90099-1
5. Hirabayashi, J.; Kasai, K.‑I. Slalom Chromatography: A New Size‑Dependent Separation
Method for DNA. Nucleic Acid Res Symp Ser 1988, 20, 67–68. DOI:
10.1021/bi00493a004
6. Hirabayashi, J.; Kasai, K.‑I. Size‑Dependent Chromatographic Separation of
Double‑Stranded DNA Which is Not Based on Gel Permeation Mode. Anal .Biochem
1989, 178, 336–341. DOI: 10.1016/0003-2697(89)90649-0
7. Hirabayashi, J.; Ito, N.; Noguchi, K.; Kasai, K.‑I. Slalom Chromatography: Size‑Dependent
Chromatographic Separation of Double‑Stranded DNA by a Hydrodynamic
Phenomenon. Biochemistry 1990, 41, 9515–9521. DOI: 10.1021/bi00493a004
8. Kasai, K.‑I. Size‑Dependent Chromatographic Separation of Nucleic Acids. J Chromatogr
B: Biomed Sci Appl 1993, 618, 203–221. DOI: https://doi.org/10.1016/03784347(93)80035-3
9. Hirabayashi, J.; Kasai, K.‑I. Effects of DNA Topology, Temperature and Solvent Viscosity
on DNA Retardation in Slalom Chromatography. J. Chromatogr. A 2000, 893, 115–122.
DOI: 10.1016/s0021-9673(00)00693-2
10. Hirabayashi, J.; Kasai, K.‑I. Applied Slalom Chromatography Improved DNA Separation
by the Use of Columns Developed for Reversed‑Phase Chromatography. J Chromatogr
A 1996, 722, 135–142. DOI: 10.1016/0021-9673(95)00505-6
11. Peyrin, E.; Guillaume, Y.; Grosset, C.; et al. Mobile‑Phase Viscosity Dependence on DNA
Separation in Slalom Chromatography. J Chromatogr A 2000, 886, 1–7. DOI:
10.1016/S0021-9673(00)00510-0
12. Peyrin, E.; Guillaume, Y. C.; Garrel, C.; et al. Effect of Temperature on DNA Fractionation
in Slalom Chromatography. Talanta 2000, 52, 1105–1110. DOI: 10.1016/s0039-
9140(00)00482-3
13. Peyrin, E.; Caron, C.; Garrel, C.; et al. DNA Migration Regimes in Hydrodynamic
Chromatography and Slalom Chromatography: Evidence for a Transition. Talanta 2001,
55, 291–296. DOI: 10.1016/S0039-9140(01)00419-2
14. Perrin, F. X.; Masuyer, C.; Truong, T. T.; et al. Supercoiled Circular DNA and Protein
Retention in Non‑equilibrium Chromatography. Temperature and Velocity Dependence:
Testimony of a Transition. J Chromatogr A 2002, 950, 281–285. DOI: 10.1016/s00219673(02)00023-7
15. Guillaume, Y.; Andre, C.; Matoga, M.; et al. Supercoiled Circular DNA Retention in
Non‑equilibrium Chromatography: Viscosity and Velocity Dependence — Behavior with
Proteins. J Chromatogr Sci. 2003, 41, 133–136. DOI: 10.1093/CHROMSCI/41.3.133
16. Andre, C.; Guillaume, Y. Novel Strategy in Slalom Chromatography for Studying Both
the Protein Reptation Mechanism and the Compacting Agent Effect to Improve
Oligonucleotide Separation. Chromatographia 2004, 59, 487–492. DOI:
10.1365/s10337-004-0242-8
17. Peyrin, E.; Guillaume, Y. C.; Villet, A.; Favier, A. Mechanism of DNA Hydrodynamic
Separation in Chromatography. Anal Chem 2000, 72, 853–857. DOI: 10.1021/ac990841s
18. Guillaume, Y. C.; Peyrin, E.; Thomassin, M.; et al. Column Efficiency and Separation of
DNA Fragments Using Slalom Chromatography: Hydrodynamic Study and Fractal
Considerations. Anal Chem 2000, 72, 4846–4852. DOI: 10.1021/ac000281t
19. Guillaume, Y. C.; Perrin, F. X.; Guinchard, C.; et al. Separation in Slalom Chromatography:
Stretching and Velocity Dependence. Anal Chem 2002, 74, 1217–1222. DOI:
10.1021/ac010852x
20. Gritti, F. Retention Mechanism in Slalom Chromatography: Perspectives on the
Characterization of Large DNA and RNA Biopolymers in Cell and Gene Therapy. J
Chromatogr A 2025, 1743, 465691. DOI: 10.1016/j.chroma.2025.465691
21. Dias, R. P. Size Fractionation by Slalom Chromatography and Hydrodynamic
Chromatography. Recent Patents Eng. 2008, 2, 95–103. DOI:
10.2174/187221208784705251
22. Li, J.; Peng, L.; Geng, X. Two Dynamic Chromatographies and their Application. Chin J
Anal Chem 2009, 37 (7), 1082–1087. DOI: 10.1016/S1872-2040(08)60118-8
23. Khanvilkar, V.; Chitnis, A.; Shirode, A.; Kadam, V. Slalom Chromatography: An Overview.
Int J Pharm Res & Chem 2012, 2, 66–76. DOI:
https://doi.org/10.1016/j.chroma.2025.465691
24. Diogo, M.; Queiroz, J.; Prazere, D. Chromatography of Plasmid DNA. J Chromatogr A
2005, 1069, 3–22. DOI: 10.1016/j.chroma.2004.09.050
25. Ghanem, A.; Healey, R.; Adly, G. Current Trends in Separation of Plasmid DNA Vaccines:
A Review. Anal Chim Acta 2013, 760, 1–15. DOI:10.1016/j.aca.2012.11.006
26. Camperi, J.; Lippold, S.; Ayalew, L.; et al. Comprehensive Impurity Profiling of mRNA:
Evaluating Current Technologies and Advanced Analytical Techniques. Anal Chem 2024,
96, 3886–3897. DOI: 10.1021/acs.analchem.3c05539
27. Delano, M.; Walter, T.; Lauber, M.; et al. Using Hybrid Organic‑Inorganic Surface
Technology to Mitigate Analyte Interactions with Metal Surfaces in UHPLC. Anal Chem
2021, 93, 5773–5781. DOI: 10.1021/acs.analchem.0c05203
28. Gilar, M.; Delano, M.; Gritti, F. Mitigation of Analyte Loss on Metal Surfaces in Liquid
Chromatography. J Chromatogr A 2021, 1650, 462247. DOI:
10.1016/j.chroma.2021.462247
29. Fekete, S.; Kizekai, L.; Sarisozen, Y. T.; et al. Investigating the Secondary Interactions of
Packing Materials for Size‑Exclusion Chromatography of Therapeutic Proteins. J
Chromatogr A 2022, 1676, 463262. DOI: 10.1016/j.chroma.2022.463262
30. Fekete, S.; Delano, M.; Harrison, A.; et al. Size Exclusion and Ion Exchange
Chromatographic Hardware Modified with a Hydrophilic Hybrid Surface. Anal Chem
2022, 94, 3360–3367. DOI: 10.1021/acs.analchem.1c05466
31. Gritti, F.; Wyndham, K. Retention Mechanism in Combined Hydrodynamic and Slalom
Chromatography for Analyzing Large Nucleic Acid Biopolymers Relevant to Cell and
Gene Therapies. J Chromatogr A 2024, 1730, 465075. DOI:
10.1016/j.chroma.2024.465075
32. Gritti, F. Theoretical Predictions to Facilitate the Method Development in Slalom
Chromatography for the Separation of Large DNA Molecules. J Chromatogr A 2024,
1736, 465379. DOI: 10.1016/j.chroma.2024.465379
33. Gritti, F. Ultra-High Pressure Slalom Chromatography: Application to the
Characterization of Large DNA and RNA Samples Relevant in Cell and Gene Therapy. J.
Chromatogr. A 2024, 1738, 465487. DOI: https://doi.org/10.1016/j.chroma.2024.465487
34. Bustamente, C.; Marko, J.; Siggia, E.; Smith, S. Entropic elasticity of 𝜆-phage DNA.
Science 1994, 265, 1599–1600. DOI: 10.1126/science.8079175
35. Marko, J.; Siggia, E. Stretching DNA. Macromolecules 1995, 28, 8759–8770. DOI:
10.1021/ma00130a008
36. Bouchiat, C.; Wang, M.; Allemand, J.; Strick, T.; Block, S.; Croquette, V. Estimating the
persistence length of a worm-like chain molecule from force-extension measurements.
Biophys. J. 1999, 76, 409–413. DOI: https://doi.org/10.1016/S0006-3495(99)77207-3
37. Gritti, F.; Sawyer, K.; Vaishnav, J.; Adapelli, B.; Lauber, M.; Wyndham, K. Retention and
efficiency of a novel slalom chromatography column: an alternative to agarose gel
electrophoresis for DNA separation. J. Chromatogr. A 2025, 1761, 466293. DOI:
10.1016/j.chroma.2025.466293
38. Strom, O.; Beech, J.; Tegenfeldt, J. High-throughput separation of long DNA in
deterministic lateral displacement arrays. Micromachines 2022, 13, 1754. DOI:
https://doi.org/10.3390/mi13101754
39. Hays, J.B.; Magar, M.E.; Zimm, B.H. Persistence length of DNA. Biopolymers 1969, 8,
531–536. DOI: https://doi.org/10.1002/bip.1969.360080410
Supplementary Figures
Figure S1a. Effect of average hydrostatic pressure (P) along a chromatographic column on
the retention of dsDNA fragments (HindIII λ-DNA digest). A very first research prototype
column (4.6 mm i.d. x 300 mm) was packed with polydisperse 3.0 μm non porous Glantreo
silica particles. Mobile phase: 1xPBS, pH 7.4. Flow rate: 0.2 mL/min. Room temperature.
Post column, 40 mm i.d. restrictor capillaries were connected in series to increase pressure:
1825 psi (no restrictor), 7040 psi (one 30 inch capillary added), 11,140 psi (one 30 inch
and one 24 inch capillary added), and 14,200 psi (one 30 inch, one 24 inch, and one 16
inch capillary added). Fact 1 ➔ DNA retention is pressure independent. Effect of average
hydrostatic pressure (P) along a chromatographic column on the retention of dsDNA
fragments (HindIII λ-DNA digest). A very first research prototype column (4.6 mm i.d. x
300 mm) was packed with polydisperse 3.0 mm non porous Glantreo silica particles.
Mobile phase: 1xPBS, pH 7.4. Flow rate: 0.2 mL/min. Room temperature. Post column, 40
mm i.d. restrictor capillaries were connected in series to increase pressure: 1825 psi (no
restrictor), 7040 psi (one30 inch capillary added), 11,140 psi (one 30 inch and one 24 inch
capillary added), and 14,200 psi (one 30 inch, one 24 inch, and one 16 inch capillary
added). Fact 1 ➔ DNA retention is pressure independent.
Figure S1b. Effect of the temperature (25–90 oC) on the retention of dsDNA fragments
(HindIII λ-DNA digest). A research prototype column (4.6 mm i.d. x 300 mm) was packed
with monosized 2.6 μm non-porous silica particles. Mobile phase: 1xPBS, pH 7.4, with
90:10 (v/v) acetonitrile. Flow rate: 1.0 mL/min. Fact 2 ➔ DNA retention decreases
progressively with increasing temperature.
Search...
Figure S1c. Effect of mobile phase viscosity (0.9–5.2 cP, corresponding to a sucrose mass
Choose
Topics fragments (HindIII λ-DNA digest). A
fraction of 0–40%) on the retention
of dsDNA
research prototype column (4.6 mm i.d. x 300 mm) was packed with monosized 2.6 μm
non-porous silica particles. Mobile phase: 1xPBS, pH 7.4. Flow rate: 0.4 mL/min. Room
temperature. Fact 3 ➔ DNA retention increases progressively with increasing viscosity of
the mobile phase.
Figure S1d. Effect of flow rate (0.15–1.1 mL/min, for 0.001xPBS-1xPBS) on the retention of
dsDNA fragments (BstEII λ-DNA digest). The commercialized 250 Å. BEH GTxResolve
Slalom Column (4.6 mm i.d. x 300 mm, 2.5-μm) was used. Mobile phase: 10xTAE, pH 8.0.
Room temperature. Fact 4 ➔ DNA retention increases progressively with increasing flow
rate.
Figure S1e. Effect of ionic strength (0.156–156 mM, duplicate measurements) on the
retention of negatively charged dsDNA fragments (HindIII λ-DNA digest). Two research
prototype columns (4.6 mm i.d. x 300 mm) were packed with monosized, negatively
charged 2.6 μm non-porous silica particles and connected in series. Mobile phase: PBS
buffers, pH 7.4. Flow rate: 0.6 mL/min. Room temperature. Fact 5 ➔ DNA retention
increases progressively with increasing ionic strength.
Figure S1f. Effect of particle size (1.8, 2.4, 5.4, and 11.3 μm) on the retention of dsDNA
fragments (λ-DNA Monocut Mix) under constant extension/shear forces (the ratio of
interstitial linear velocity to particle diameter was kept constant) across a wide range of
average extension lengths (Lext: 0.05–13.5 μm). Four research prototype columns (4.6 mm
i.d. x 300 mm) were packed with 125 Å BEH particles of different sizes. Mobile phases: 100
mm phosphate buffer, pH 8.0. Flow rates: 0.262 mL/min (1.8 μm), 0.436 mL/min (2.4 μm),
0.921 mL/min (5.4 μm), and 2.0 mL/min (11.3 μm). Room temperature. Fact 6 ➔ DNA
retention is independent of particle size when operating at constant extension/shear force
and when the extension length exceeds the particle diameter. Fact 7 ➔ Conversely, DNA
retention decreases with increasing particle size at constant extension/shear force when
the extension length is smaller than the particle diameter, providing evidence of a mixed
HDC–SC retention mechanism.
Figure S1g. Effect of surface modification of stainless steel column hardware with hybrid
surface technology (HST) bonding on the peak shape of dsDNA fragments (HindIII λ-DNA
digest). The same 2.5 μm, 250 Å BEH particles and column dimensions (4.6 mm i.d. x 300
mm) were used. Mobile phase: 1xTAE, pH 8.0. Flow rate: 1.4 mL/min. Temperature: 40 oC.
(Top) Bare stainless-steel column hardware. (Bottom) HST-modified column hardware.
Fact 8 ➔ The integrity of DNA peak shape and retention is maintained with HST modified
stainless steel column hardware.
Figure S2. Effect of TAE buffer concentration (4–400 mM) on the retention of negatively
charged dsDNA fragments (from HindIII λ-DNA digest, BstEII λ-DNA digest, and λ-DNA
Monocut Mix) ranging in size from 1.5 to 48.5 kbp. The commercialized 2.5 μm, 250 Å
BEH GTx- Resolve Slalom Column (4.6 mm i.d. x 300 mm) was used. Mobile phases:
0.1xTAE, 1xTAE, and 10xTAE, pH 8.0. Flow rate: 1.1 mL/min. Room temperature. Again, as
in Figure S1e, DNA retention increases with increasing ionic strength.
Newsletter
Join the global community of analytical scientists who trust LCGC for insights on the
latest techniques, trends, and expert solutions in chromatography.
Subscribe Now!
Column: Column Watch
Related Content
SPONSORED
Clarus® Nova GC: One Platform. More
Capability. Less Complexity.
Discover how a unified GC platform can help laboratories
improve uptime streamline workflows and stay prepared for…
,
Learn More
,
Pits as Protectors: LC-MS/MS Unlocks Antioxidant and Anti-Browning Potential
By John Chasse
April 17th 2026
About
Advertise
Author Guidelines
Contact Us
Editorial Advisory Board
Ethics Policy
Do Not Sell My Personal Information
Privacy Policy
Permissions
Subscriptions
Terms and Conditions
Cookie settings
Contact Info
259 Prospect Plains Rd, Bldg H
Cranbury, NJ 08512
Best of the Week: Sustainable Analytical Methods, Chemometrics and Generative
AI
By John Chasse
April 17th 2026
© 2026 MJH Life Sciences®
All rights reserved.
Home
About Us
News
Contact Us
Invasive Burmese Pythons: Sentinels for PFAS Pollution in the Everglades via HPLCMS/MS Analysis
By John Chasse
April 17th 2026
The SWIM Approach Extended
By Gauthier Eppe, Kate Jones
April 17th 2026
Chromatographic Profiling of Steroid Hormone Variability Across Menstrual Cycle
and Hormonal Contraceptive Use in a Female Athlete
By John Chasse
April 16th 2026